CLEANROOM VALIDATION: Tier 1
VALIDATION: Certification to be carried out in accordance with BS/EN/ISO 14644 – 1
MAGNAHELIC CALIBRATION: All Cleanroom Magnahelic Gauges will be calibrated.
AIRFLOW VELOCITY: Airflow velocity will be measured at 5 positions on each HEPA filter.
PARTICLE COUNTS: Particle counts to be carried out in each room at bench height. Each sample will be for 1 minute, with a flow rate of 1cfm set on the particle counter.
TEMPERATURE & HUMIDITY, LUX LEVEL: Lighting, Temperature & Humidity to be measured at a minimum of 8 positions in the room.
REPORT: A detailed report will be generated showing all monitoring points, raw particle count data printouts, airflow velocities and air changes per hour calculations, together with a layout of the rooms and cabinets. A controlled environment certificate will be issued detailing the report number and classification achieved in BS/EN/ISO 14644-1.